Lens paper data are single channel and they no longer have a clear purpose #79

Open
raacampbell opened this issue Jan 21, 2025 · 1 comment

Comments

@raacampbell
Copy link
Collaborator

raacampbell commented Jan 21, 2025

The lens paper was being imaged to provide a quick proxy for assessing the overall microscope health. This was almost a subjective check: "If I can see lens paper at 15 mW then things can't be too bad". So the data were acquired with just one channel.

I think we now have a better plan: Image lens paper with all channels at some reasonably high laser power and calculate photons/pixel. Possibly repeat this at a few gains for sanity. The reported result would therefore be the variance/mean fit plus, if possible, a standard light source converted to photons at that same gain and pixel size.

details at

@raacampbell raacampbell self-assigned this Jan 21, 2025
@raacampbell raacampbell changed the title Lens paper data are single channel Lens paper data are single channel and they no longer have a clear purpose Feb 13, 2025
@raacampbell
Copy link
Collaborator Author

Ok. Lens paper now being saved over multiple channels. However, we need to:

  • Set gains of all PMTs to the same thing. Could be tricky on setups with different PMT types
  • Ensure all channels are set to be acquired.